PRODUCT

CAPILLARIES BASED GAS BLENDER FOR

GAS ANALYSERS TEST - BETACAP30

 

DESCRIPTION

    Sometime it's required the availability of many gas concentrations for the calibration of an analyser in various points of the range or in various ranges. Zero and sensitivity calibrations are not enough for assuring the instrument accuracy because the non linearity response is disregarded. The response check of an instrument in some point of the range is not only suggested by “good practice”, but requested by environmental laws and by norms for approval of components which involves combustion (engines, burners, boilers, …).

BetaCAP30 is made by 30 identical capillaries, which are individually connected to the gas to be diluted or to the diluting gas. It's then generated a mixture which may have one of 31 different concentration values, depending on the ratio of the capillaries number connected to one source and the capillaries number connected to the other source.

Each capillary will be crossed by the gas to be diluted or by the diluting gas : the diluted flow will be 30 times the single capillary flow.

To assure the matching of all the capillary flows, the pressure of the 2 gas inlets must be adjusted at the same value or at suitable values to compensate the differences in density and viscosity of the different gases. For that reason, differential pressure across capillaries is continuously measured and displayed to support the user.

A suitable adjustment of inlet pressures, allows the production of different gas flows : selecting pressures in the range 50...200 kPa, the output flow may change in the range 1...4 litres/min. Upon request, different capillary sizes may be selected to cover different ranges.

Two fine filters protects from dust the pressure regulators and capillaries : keeping clean that area is the best assurance for high accuracy and long life of the gas blender.

In addition, the mechanical assembling is rugged and shocks absorbing, to avoid misalignments or cracks and compact in size to be easily transported.

The main part of the instrument is a manifold which holds in place and create the circuit for the capillaries : they are got by cutting one long capillary tube . On the same manifold, solenoid valves with separation diaphragm are faced to address the right gas inlet to the right capillaries group. Individuals capillaries are selected by fluid-dynamic test and matched before installation, to get the best accuracy of the unit .

All of the internal parts may be inspected for cleaning or replacement, and the used materials are highly resistant to aggressive gases : Aisi 316, borosilicate glass, PVDF, PEEK, Kalrez, epoxy resin.

A microprocessor electronic circuit manages all the functions interfacing with the user through an alphanumeric display and five function keys.

Two main menus are used to calibrate the input pressure measurements and the input current signal from the instrument under test. One menu is used to define parameters like test gas bottle concentration and analyser measuring range. Two menus are ready for operating the dilutor locally or remotely from a PC via RS232.

The user inputs test gas bottles and analyser data, and then selects the number of dilution steps : theoretical and measured concentration is automatically displayed.

Through the serial interface, a PC with the suitable software (Terminal or a dedicated program) may send to the gas divider the dilution ratio and receive the instrument measuring signal : recording

those values for each step, the test report may be automatically drawn.

  TECHNICAL SPECIFICATIONS

Dilution ratio:

adjustable from 0 to 100% in 32 steps

Pneumatic taps inlets :

diluting gas and gas to be diluted

outlet :

diluted gas

Accuracy:

better than 0.2 % of the concentration to be diluted

Repeatability :

better than 0.1 % of the concentration to be diluted.

Gas contact. materials :

AISI 316, borosilicate glass, PVDF, PEEK, Kalrez, epoxy resin.

Analog measurements 2 differential pressures and gas analyser measurement

Size and weight :

portable version 270 x 250 h 180 mm - weight 10 kg - IP67

laboratory version rack 19” std. 3UT - dept 250 mm

Serial interface :

RS232 to download test data to a PC

Power supply :

110 / 220 Vac - 2 A - (different values ac / dc at request)

 

 

 

 